Weight:
12,362 all components
Axial Fov, Mm :
157
Cooling, Btu/Hr :
31,000 (min); no chilling required
Crystal Size, Mm :
6.2 x 6.2 x 30
Detector Rings :
24
Dispersion Fraction :
19% (2-D), 44% (3-D). 2001 NEMA performance specifications (NU 2-2001)
Display:
Two 21' tilt/swivel high-res color
Maximum Count Rate, Cps @50% Dead Time :
84 kcps @ 49 kBq/cc (2-D), 63 kcps @ 12 kBq/mL (3-D); NECR. 2001 NEMA performance specifications (NU 2-2001)
Minimum Room Size :
4.4 x 7.8 m (14.5 x 25.5 ft)
Patient Weight Limit Kg :
180 (400 lbs.) w/0.25 mm positional accuracy
Power Supply Voltage:
380-480 VAC nominal, 3-phase, 50/60 Hz
Printer :
DICOM
Relative Humidity % :
30-60%
Ring Diameter, Cm :
88.6
Storage Capacity :
36 GB raw data disk, 18 GB data disk (CT)
System Sensitivity :
2 cps/kBq (2-D), 9.3 cps/kBq (3-D)