SELLER is offering PM preventative maintenance on the INTEGRA CUSA EXCel surgical aspirator. Preventative maintenance on the INTEGRA CUSA NXT surgical aspirator and preventative maintenance on the INTEGRA CUSA EXCel+ surgical aspirator.

Send us a copy of the Integra estimate and we will beat their price. If you have previous PM documentation, plesas send them to us and we will create a documentation file on your system for you to have access to at any time.

If you have any questions, contact us and ask for CUSA Service.

Th PM on your system maximizes Functionality and our Technicians Stand Behind Every service event.
The PM service includes system re-calibration and adjustment for precise performance. Replace all PM components. Repaired with only genuine Integra original equipment manufacturer (OEM) parts.


**************************************************
ALERTS or ERROR CONDITION

ALERT-Footswitch
CAUSES-Footswitch cable not properly connected to console
RECOMMENDED ACTION-Connect the footswitch to the console
(rear panel). Turn the connector until you feel it lock into place. If the alert persists after you connect the
footswitch to the console, call SELLER for assistance.

ALERT-Handpiece
CAUSES-Handpiece cable not properly connected to console
RECOMMENDED ACTION-Connect the handpiece to the console. Twist the handpiece connector clockwise until it locks into place, then verify that the yellow dots align. If the alert persists after you connect the handpiece to the console, call Integra for assistance.

CAUSES-Using a CUSA 200 system 23 kHz handpiece (black connector)
RECOMMENDED ACTION-Connect a CUSA EXcel 23 kHz handpiece (green connector) or 36 kHz handpiece (blue connector).

ALERT-Cooling Water—continuously on
CAUSES-Not enough distilled water in the cooling water reservoir
RECOMMENDED ACTION-Add distilled water to the cooling water reservoir to the fill line.

CAUSES-Cooling water reservoir not properly seated
RECOMMENDED ACTION-Re-seat the cooling water reservoir in the console.

CAUSES-A pinch or kink in the handpiece cable
RECOMMENDED ACTION-Remove the pinch or kink in the cable.

CAUSES-Handpiece cable not properly connected to console
RECOMMENDED ACTION-Connect the handpiece to the console.
Twist the handpiece connector clockwise until it locks into place, then verify that the yellow dots align.

CAUSES-Damaged, misaligned, or missing o-rings in the handpiece connector
RECOMMENDED ACTION-Connect a new handpiece to the CUSA EXcel system, and test it for proper function. If the alert persists after you connect the new handpiece to the console, call Integra for assistance.